Ultrasound System Cleaning And Disinfection
This manual is provided as an emergency process in response to the outbreak of coronavirus disease 2019 (COVID-19).
The contents differ slightly from the existing manuals included with the Ultrasound system as this process is provided specifically for the current outbreak to simplify and clarify the process on site and prevent virus transmission.
Please refer to the websites below for more information.
World Health Organization (WHO)
• Infection prevention and control during health care when novel coronavirus (nCoV) infection is suspected
Centre for Disease Control (CDC) USA
• Interim Guidance for Healthcare Facilities: Preparing for Community Transmission of COVID-19 in the United States
• Implement Environmental Infection Control
Environmental Protection Agency (EPA) USA
• List N: Disinfectants for Use Against SARS-CoV-2
National Institute of Infectious Diseases, Japan
• Infection control for COVID-19
Canon cleaning, disinfection, and sterilization Manuals
• Ultrasound System (Select Model Name for system manual)
• Ultrasound transducers and transducer accessories
Note1: Sodium hypochlorite is low level disinfection. Use high-level disinfection for equipment that comes into contact with mucous membranes.
Note2: Ethanol (76.9 to 81.4 vol % at 15°C) or Isopropyl alcohol (70 vol %) can be used in place of sodium hypochlorite.
Note3: Refer to the manual for details.
1: Cleaning
2: Disinfection
Disinfection should be in compliance with the instructions of the infection control department
1: Cleaning
Required : Personal Protection Equipment (PPE) such as a Face Mask, Eye Protection, Apron and Gloves, cleaning solution or cleaning wipes, purified water, clean soft cloth or gauze, non-abrasive single-use sponge
2: Disinfection
Before disinfection, the transducer must be cleaned and dried.
Note that local regulations may require that the transducer be disinfected before sterilization.
Required : Personal Protection Equipment (PPE) such as a Face Mask, Eye Protection, Apron and Sterile Gloves, disinfectant, sterile water or deionized water, sterile soft cloth or gauze
●Have the disinfectant chemicals been tested for efficacy against SARS-CoV-2?
Canon Medical has not tested the chemicals listed for efficacy , the information provided has been referenced from the United States Environmental Protection Agency (EPA) and who have released a list of disinfectants that claim effectiveness for use against SARS-CoV-2. Sodium hypochlorite, ethanol and isopropyl alcohol from that list are recommended in the guidelines.
●Can the Ultrasound system fan spread the virus?
Most electronic devices including ultrasound systems have essential fans to cool the components. There is no clear data at this time regarding the possible spread of COVID-19 by electronic equipment fans.
●Are air filters in Ultrasound systems effective at removing viruses?
The filters in electronic devices are designed to prevent dust & lint entering the device, this is also true for Ultrasound systems. The filters are not intended to block specific pathogens from entering the system.