Scan Mode | Helical |
Collimation | 0.5mm x 40 |
kVp | 120 |
mAs | 240 |
Rotation Time | 0.75s |
Scan Range | 140.0mm |
Reconstruction | AiCE Brain LCD |
CTDIvol | 57.0mGy |
DLP | 915.8mGy·cm |
Effective Dose | 5.4mSv |
k-factor | 0.0021 (AAPM Report 96) |
Auto subtraction demonstrates intracranial aneurysms associated with AVM.
View Scan Parameters (Non-Contrast)Scan Mode | Ultra-Helical |
Collimation | 0.5mm x 80 |
kVp | 120 |
mAs | 165 |
Rotation Time | 0.75s |
Scan Range | 225.0mm |
Reconstruction | AiCE |
CTDIvol | 51.5mGy |
DLP | 953.1mGy·cm |
Effective Dose | 2.00mSv |
k-factor | 0.0021 (AAPM Report 96) |
Scan Mode | Ultra-Helical |
Collimation | 0.5mm x 80 |
kVp | 120 |
mAs | 165 |
Rotation Time | 0.75s |
Scan Range | 225.0mm |
Reconstruction | AiCE |
CTDIvol | 35.2mGy |
DLP | 698.8mGy·cm |
Effective Dose | 1.46mSv |
k-factor | 0.0021 (AAPM Report 96) |
CTA with SURESubtraction demonstrates a right ICA occlusion.
View Scan Parameters (Non-Contrast)Scan Mode | Ultra-Helical |
Collimation | 0.5mm x 80 |
kVp | 120 |
mAs | 150 |
Rotation Time | 0.75s |
Scan Range | 164.0mm |
Reconstruction | AiCE |
CTDIvol | 42.1mGy |
DLP | 871.4mGy·cm |
Effective Dose | 1.83mSv |
k-factor | 0.0021 (AAPM Report 96) |
Scan Mode | Ultra-Helical |
Collimation | 0.5mm x 80 |
kVp | 120 |
mAs | 187 |
Rotation Time | 0.75s |
Scan Range | 164.0mm |
Reconstruction | AiCE |
CTDIvol | 52.7mGy |
DLP | 1089.3mGy·cm |
Effective Dose | 2.28mSv |
k-factor | 0.0021 (AAPM Report 96) |
AiCE DLR enabling high diagnostic confidence in evaluation of small bony structures in IAC, through significant noise reduction while maintaining high-resolution.
View Scan ParametersScan Mode | Ultra-Helical |
Collimation | 0.5mm x 40 |
kVp | 120 |
mAs | 150 |
Rotation Time | 0.5s |
Scan Range | 57.0mm |
Reconstruction | AiCE Inner Ear |
CTDIvol | 35.6mGy |
DLP | 181.3mGy·cm |
Effective Dose | 0.4mSv |
k-factor | 0.0021 (AAPM Report 96) |
SURESubtraction automatically generates images with subtracted bones improving the visualisation of the neurovascular anatomy.
View Scan Parameters (Non-Contrast)Scan Mode | Ultra-Helical |
Collimation | 0.5mm x 80 |
kVp | 100 |
mAs | SUREExposure |
Rotation Time | 0.5s |
Scan Range | 350.0 mm |
Reconstruction | AiCE |
CTDIvol | 3.9mGy |
DLP | 152.7mGy·cm |
Effective Dose | 0.47mSv |
k-factor | 0.0031 (AAPM Report 96) |
Scan Mode | Ultra-Helical |
Collimation | 0.5mm x 80 |
kVp | 100 |
mAs | SUREExposure |
Rotation Time | 0.5s |
Scan Range | 350.0 mm |
Reconstruction | AiCE |
CTDIvol | 3.9mGy |
DLP | 152.7mGy·cm |
Effective Dose | 0.80mSv |
k-factor | 0.0031 (AAPM Report 96) |
Contrast enhanced CT of the head with AIDR 3D demonstrates a brain hemorrhage.
View Scan ParametersScan Mode | Ultra-Helical |
Collimation | 0.5mm x 80 |
kVp | 120 |
mAs | 225 |
Rotation Time | 0.5s |
Scan Range | 162.0 mm |
Reconstruction | AIDR 3D |
CTDlvol | 27.2mGy |
DLP | 561.0mGy·cm |
Effective Dose | 1.17mSv |
k-factor | 0.0021 (AAPM Report 96) |
Iodine maps,110 Equiv. kV Enhanced images demonstrate tumor enhancement, the VNC images show bone destruction.
View Scan ParametersScan Mode | DE-Helical |
Collimation | 0.5mm x 80 |
kVp | 80/135 |
mAs | 100/570 |
Rotation Time | 0.5s |
Scan Range | 310.0mm |
Reconstruction | AIDR 3D |
CTDIvol | 21.4mGy |
DLP | 688.8mGy·cm |
Effective Dose | 1.14mSv |
k-factor | 0.0021 (AAPM Report 96) |